Comparative calibration vs. fixed point calibration

In temperature calibration technology, you can find two ways of calibration ? the comparative calibration and the fixed point calibration. The technique of calibration to be utilized will be defined through the demands on the accuracy and the type of test instrument. The comparative method represents probably the most widely used method of calibration.
So how do the comparative calibration and fixed point calibration methods differ? Utilizing the exemplory case of calibrating a resistance thermometer, this blog article explains both methods and thus tries to mention to the reader a better understanding.
The comparative calibration
Because the name already implies, this type of calibration is carried out by comparing the resistance thermometer being tested with a higher-quality reference thermometer. The higher-quality reference should thus exhibit an accuracy of 3 or 4 times higher than the thermometer being tested. Both thermometers are placed into a single homogenous temperature source. When placing the thermometers, care ought to be taken to ensure a little distance between them and that the sensitive points (meaning the measuring points) are in the same height. As a temperature source, liquid baths or dry-well calibrators are usually used.
The fixed point calibration
The fixed point calibration, as opposed to the comparative calibration, is completed contrary to the defined triple, freezing and melting points from the International Temperature Scale of 1990 (ITS-90). The fixed points define temperatures at which the physical state of substances alter. The most important fixed point of the ITS-90 may be the triple point of water. At a temperature of 0.010 �C the physical state of high-purity water is continually changing between steam, ice and liquid water. The levels of the physical states always stay in proportion. Generally, only national laboratories or laboratories qualifying for very small measurement uncertainties work with these fixed-point cells, as they are costly and specialised only for this application. Backbone (platinum resistance thermometers), which will be used as reference thermometers, are calibrated using the fixed point method. Youthful is inserted into a closed, or also open fixed-point cell, which holds the temperature constant over a long period.
Example: fixed points from -189 to 660 �C in accordance with ITS-90:
Fixed point
Temperature
Triple point of argon
-189,3442 �C
Triple point of mercury
-38,8344 �C
Triple point of water
0,01 �C
Melting point of gallium
29,7646 �C
Freezing point of indium
156,5985 �C
Freezing point of tin
231,928 �C
Freezing point of zinc
419,527 �C
Freezing point of aluminium
660,323 �C
Note
More info on our Services ?Temperature calibration? can be found on the WIKA website.

Leave a Comment